Herunterladen Inhalt Inhalt Diese Seite drucken

Assembly Iris Led 10 W; Load Data; Assembly Arm System - Derungs D med IRIS LED 10 Montage- Und Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Connect luminaire to mains
Push cover and end ring over the ceiling bracket
and tighten them (0.5Nm)
Continue with luminaire head assembly, see
Chapter 7
5

ASSEMBLY IRIS LED 10 W

5.1

Load data

Bending moment M
B
Vertical weight F
G
5.2

Assembly arm system

DANGER
Assembly by qualified personnel
Assembly must be carried out by qualified
personnel only. Lack of appropriate knowledge
could be life-threatening.
DANGER
Life-threatening danger from falling luminaire
The ceiling must be made of solid concrete to
guarantee a secure hold.
Electric shocks are life threatening
Electric shocks are life-threatening
CAUTION
Take account of the wall bracket position
160Nm
108N
DANGER
It must be possible to separate the light from
the mains by an external all-pole switch
(not included).
DANGER
Voltage supply from the power connection
must be protected on all poles by means of an
overcurrent release (according to national
installation requirements; it is not included in
the scope of supply).
The wall bracket must be aligned vertically, as
shown in the image.
Failure to comply with correct alignment will
compromise mechanical safety.
Make minimum 4 opposited drill hole markings
(Ø13mm), note the position of the Ø60mm drill
hole for electrical connections
Check clearances
Drill the bore holes, then blow them out using
bellows