Safety Instructions; Intended Use; User Profiles; Warning Levels - Derungs D med IRIS LED 10 Montage- Und Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ENG
1

SAFETY INSTRUCTIONS

1.1

Intended use

med®
The luminaire D
IRIS LED is an examination light.
Its intended use is to illuminate the body of the patient
to support diagnosis or treatment. An interruption of
the diagnosis or treatment caused by a light failure is
always possible without posing any risk for the patient.
The light cannot be used in operating rooms.
1.2

User profiles

Medical professional
A medical professional is any person who has
completed medical training and is working in his or her
professional field.
Cleaning professional
A cleaning professional is trained in national and job-
related hygiene regulations.
Qualified electrician
A qualified electrician is trained in the fields of
electronics and electrical engineering and is familiar
with the relevant standards and regulations.
Qualified professional
A qualified professional is capable of the assembly and
disassembly of the luminaire thanks to professional
training, knowledge and experience and knowledge of
the regulations.
1.3
Safety instructions
Operation by a medical professional
The instructions form part of the product. They
must be stored and made accessible to all
subsequent users.
Any work on the luminaire (including repairs) must
be carried out by qualified electricians only.
Assembly must be performed by a qualified
professional only.
The luminaire must not be altered or manipulated
in any way. Only approved original parts must be
used. Any use other than the intended use with
original parts may give rise to other technical
values and life-threatening danger.
Operation
in
potentially
prohibited. The luminaire power supply is a
potential ignition source.
The luminaire must only be operated in dry, dust-
free rooms.
The luminaire must not be left switched on without
supervision.
For luminaires in protection class I, it is essential
for the earthing conductor to be connected to the
luminaire housing.
Do not use any luminaire that is damaged.
Defective cables and a defective handle are also
potential hazards. Do not place the cable close to
any heat source or on sharp edges.
Damage to the eyes. Never look directly into the
light source.
Replace
any
glass
operating the luminaire again.
In order to prevent electric shocks, only connect
the luminaire to a power supply with an earth
conductor.
Never place additional loads on the luminaire head
or the arm.
24
explosive
areas
that is
damaged
before
The luminaire must not be covered by a cloth or
any similar item while it is in operation.
The ventilation openings (where they exist) must
be free whenever the luminaire is in operation.
The luminaire must not be operated near to
external heat sources that exceed the maximum
ambient temperature of the luminaire.
The luminaire must not be used in environmental
conditions other than those for which it is
intended.
The luminaire must only be used for the intended
use described in this document.
The manufacturer cannot be held responsible for
any injury or damage that is caused as a result of
any use other than the intended use or of any
failure to comply with safety instructions and
warnings.
The IRIS LED 10 F must be secured while being
transported within the clinic.
1.4

Warning levels

DANGER
Warnings of hazards that can lead to death or
serious injury if measures are disregarded.
WARNING
Warnings of hazards that can lead to injuries if
measures are disregarded.
CAUTION
Warnings of hazards that can lead to damage to
property if measures are disregarded.
1.5

Specific mounting instructions

IRIS LED 10 C
Mounting material is not included.
The ceiling mounting must only be fixed to ceilings
that have a concrete stability class B25 (C20/25) or
higher.
Solid ceiling reinforcement parts must not come
is
into contact with the ceiling mounting. If there is
any doubt, an authorized professional must
confirm that mounting is possible on the mounting
base being used.
Borings must be made by specialists in compliance
with the boring tolerances for the reinforcement
anchor
manufacturer.
Screws must be tightened carefully using a torque
key that complies with the instructions of the
reinforcement anchor manufacturer.
IRIS LED 10 W
Mounting material is not included.
The luminaire must only be mounted on walls that
can guarantee a secure hold. Specialist staff will
find details of the requirements in Chapter 6.1
(Load data).
IRIS LED Duo 10/10 C
and
that
have
been
approved
by
the

Werbung