Herunterladen Inhalt Inhalt Diese Seite drucken

Assembly Iris Led Duo 10/10 C; Load Data; Assembly Arm System - Derungs D med IRIS LED 10 Montage- Und Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ENG
Connect luminaire to mains
Push cover and end ring over the ceiling bracket
and tighten them (0.5Nm)
Continue with luminaire head assembly, see
Chapter 7
4
ASSEMBLY
IRIS LED Duo 10/10 C
4.1

Load data

Bending moment M
B
Vertical weight F
G
4.2

Assembly arm system

DANGER
Assembly by qualified personnel
Assembly must be carried out by qualified
personnel only. Lack of appropriate knowledge
could be life-threatening.
DANGER
Life-threatening danger from falling luminaire
The ceiling must be made of solid concrete to
guarantee a secure hold.
28
Electric shocks are life threatening
Electric shocks are life-threatening
Pay attention to the minimum distance of the
ceiling tube
Saw ceiling tube to the desired length at the top
end using a metal saw and deburr.
Remove fastening screw.
CAUTION
Prior to assembly, check the reach and position of
the luminaire in the direction of the desired field.
90Nm
206N
DANGER
It must be possible to separate the light from
the mains by an external all-pole switch
(not included).
DANGER
Voltage supply from the power connection
must be protected on all poles by means of an
overcurrent release (according to national
installation requirements; it is not included in
the scope of supply).

Werbung

Inhaltsverzeichnis
loading