Herunterladen Inhalt Inhalt Diese Seite drucken

Derungs D med IRIS LED 10 Montage- Und Gebrauchsanweisung Seite 32

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ENG
Hold the wall bearing onto the wall and strike
the wall anchor with a hammer
Tighten fastening in line with manufacturer
data
WARNING
The
light
will
fall
incorrectly
The threaded pin must be screwed in fully and
correctly.
Perform this assembly step on a flat
surface:
Lay arm 1 down flat
Pull cable (from the arm) through wall angle 2
Insert slide pad 3 into the groove at the top
Push wall angle 2 (with boring 4 aligned
upward) and arm 1 together
Secure slide pad 2 by screwing in the threaded
pin 5 (up to the stop)
Release threaded pin 5 by a ¼ turn maximum
and check function
Also check:
Threaded pin must now be flush
Remove fastening screw.
32
if
fastening
is
done
Insert wall angle 1 (with closure ring 2 and
hood 3 in position) into the wall bracket 4 and
simultaneously pull the cable through the
rectangular cutout
Wall angle in vertical alignment, mount with M8
self-locking bolt and nut and tighten 1 (20 Nm)
Tighten all 4 threaded pins 2 (5 Nm)
DANGER
Mortal danger from electric shock
Disconnect from the power supply and secure
against unintended switching on before starting
work on the power connection.
Make power connection
Connect luminaire to mains