Uric Acid Test:
Message
< 3 mg/dL (0.179 mmol/L)
> 20 mg/dL (1.190 mmol/L)
Error Message
Error
Cause
Message
The batteries cannot
provide enough power
for a test.
Strip has been used.
Problem in operation.
What it Means
What To Do
Replace the battery
immediately and reset
date and time on the
meter setting.
Repeat the test with a
new strip.
Review the instructions
and repeat the test with
a new strip. If problem
persists, contact the
local customer service
for assistance.
You may have removed
the strip after applying
blood, or insufficient
blood volume.
Ambient temperature
is below the system's
operation range.
Ambient temperature
is above the system's
operation range.
Appears when the test
strips are expired.
(This only applies
to ß-ketone, total
cholesterol and uric
acid test strips)
EN-20
Review the instructions
and repeat test with a
new test strip.
System operational
range is 8°C to 45°C
(46.4°F to 113°F).
Repeat the test after
the device and test strip
have reached the above
temperature.
Discard the expired
test strips. Use a new
test strip which doesn't
exceed the expiry date.