Performing A Test

1. Insert the test strip into the test slot of the device. Wait for
the device to display the test strip "
".
2. Press
or
to adjust the measuring mode, and press
MAIN button to confirm it.
• General Tests (
) - any time of day without regard to
time since the last meal.
• AC (
) - no food intake for at least 8 hours.
• PC (
) - 2 hours after a meal.
• QC (
) - testing with control solution.
3. Obtain a blood sample.
Use the pre-set lancing device to puncture the desired
site. Wipe off the first appeared drop of blood with a clean
cotton swab. Gently squeeze the punctured area to obtain
another drop of blood. Be careful NOT to smear the blood
sample.
The blood sample size of each test should be at least,
• blood glucose test: 0.5μL
• blood glucose/ hematocrit/ hemoglobin test: 0.5μL
" and blood drop "
EN-11
• β-ketone test: 0.8μL
• total cholesterol test: 3.0μL
• uric acid test: 1.0μL
4. Apply the blood sample.
Move your finger to meet the absorbent hole of the test
strip and the drop will be automatically be drawn onto
the test strip. Remove your finger until the confirmation
window is filled. The meter begins to count down. Do not
remove your finger until you hear a beep sound.
5. Read your result.
The results of your blood glucose test will appear after
the meter counts down to 0. The results will be stored
automatically in the meter memory.