Table 7 Warning list for conductivity sensors (continued)
Warning
Description
CAL OVERDUE
The Cal Reminder time
has expired
NOT
The sensor has not been
CALIBRATED
calibrated
REPLACE
The sensor has been in
SENSOR
operation > 365 days
CAL IN
A calibration was started
PROGRESS
but not completed
OUTPUTS ON
During calibration, the
HOLD
outputs were set to hold for
a selected time.
WRONG LINEAR
The user-defined linear
TC
temperature compensation
is out of range
WRONG TC
The user-defined
TABLE
temperature compensation
table is out of range
Event list for sensors
The event list shows current activities such as configuration changes,
alarms, warning conditions, etc. To show the events, press the MENU
key and select Sensor Diag, [Select Sensor], Event List. A list of possible
events is shown in
Table
8. Previous events are recorded in the event
18 English
Resolution
Calibrate the sensor.
Calibrate the sensor.
Calibrate the sensor with a
reference solution and reset
the sensor days. Refer to
Sensor diagnostic and test
menu
on page 16. If the
calibration fails, call technical
support.
Return to calibration.
The outputs will become active
after the selected time period.
The value must be between
0 and 4%/°C; 0 to 200 °C.
The temperature is above or
below the temperature range
defined by the table.
log, which can be downloaded from the controller. Refer to the controller
documentation for data retrieval options.
Table 8 Event list for conductivity sensors
Event
Description
CAL READY
The sensor is ready for calibration
CAL OK
The current calibration is good
TIME EXPIRED
The stabilization time during calibration expired
CAL FAIL
The calibration failed
CAL HIGH
The calibration value is above the upper limit
K OUTRANGE
The cell constant K is out of range for the current
calibration
UNSTABLE
The reading during calibration was unstable
CHANGE IN CONFIG float The configuration was changed—floating point type
CHANGE IN CONFIG text
The configuration was changed—text type
CHANGE IN CONFIG int
The configuration was changed—integer value type
RESET CONFIG
The configuration was reset to the default options
POWER ON EVENT
The power was turned on
ADC FAILURE
The ADC conversion failed (hardware failure)
FLASH ERASE
The external serial flash memory erase occurred
TEMPERATURE
The temperature is out of range (-20 to 200 °C)
SAMPLE CAL START
Start of calibration for conductivity
SAMPLE CAL END
End of calibration for conductivity
ZERO CAL START
Start of zero calibration
ZERO CAL END
End of zero calibration