Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting - Hach Polymetron 8310 Bedienungsanleitung

Conductivity sensors
Vorschau ausblenden Andere Handbücher für Polymetron 8310:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 20

Troubleshooting

Intermittent data
During calibration, data is not sent to the datalog. Thus, the datalog can
have areas where the data is intermittent.
Test the conductivity sensor
If a calibration fails, first complete the maintenance procedures in
Maintenance
on page 15.
1. Disconnect the sensor wires.
2. Use an ohmmeter to test the resistance between the sensor wires as
shown in
Table
4.
Note: Be sure that the ohmmeter is set to its highest range for all infinite (open
circuit) resistance readings.
Table 4 Conductivity resistance measurements
Measurement points
Between blue and white wires
Between red wire and sensor body
Between black wire and inner electrode
Between black and red wires
Between black and white wires
Between red and white wires
Between red and inner shield wires
Between black and inner shield wires
Between white and inner shield wires
Between outer and inner shield wires
If one or more of the measurements is incorrect, call technical support.
Supply technical support with the serial number of the sensor and the
resistance values measured.
16 English
Resistance
1089–1106 ohms at 23–27 °C
Less than 5 ohms
Less than 5 ohms
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Infinite (open circuit)
Sensor diagnostic and test menu
The sensor diagnostic and test menu shows current and historical
information about the instrument. Refer to
diagnostic and test menu, push the MENU key and select Sensor Setup,
[Select Sensor], DIAG/TEST.
Table 5 Sensor DIAG/TEST menu
Option
Description
SENSOR INFORMATION Shows the name and serial number that was entered
by the user.
CARD INFORMATION
Shows the version and the serial number for the
sensor module.
CAL DAYS
Shows the number of days since the last calibration.
CAL HISTORY
Shows a list of the calibrations and the details for
each calibration.
RESET CAL HISTORY
Service use only. Resets the calibration history for
the sensor. All previous calibration data is lost.
SENSOR SIGNALS
Shows the current sensor signal and span in µS/cm.
SENSOR DAYS
Shows the number of days that the sensor has been
in operation.
RESET SENSOR DAYS
Resets the Sensor Days counter.
Error list
When an error occurs, the reading on the measurement screen flashes
and all outputs are held when specified in the controller menu. To show
Table
5. To access the sensor

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis