Service Notes (Continued)
5. Overall surface inspection - Review the overall surface of the Skull Clamp to check for any
cracks or fractures of the components. This is a molded composite plastic that can suffer cracks
or breakage when subjected to rough handling or being dropped. In the event that the Skull
Clamp or a component of the Skull Clamp is dropped on the floor or sharply hit, it is highly
recommended that all components be returned for inspection by your authorized MAYFIELD
Repair Center.
Service and Repair
For service and repairs outside the United States, contact your local authorized Integra representative.
Within the United States, send all instruments for service or repair to:
Integra LifeSciences Corporation
4900 Charlemar Drive, Building A
Cincinnati, Ohio 45227
(Always include the purchase order number and a written description of the problem)
Or phone: 877-444-1114
U.S. Patent 7.507.244; additional patent(s) pending
Integra and the Integra logo are registered trademarks of Integra LifeSciences Corporation in the
United States and/or other countries. MAYFIELD is a registered trademark of SM USA Inc. and is used
by Integra under license. Endozime is a trademark of Ruhof Corporation. Renu-Klenz is a trademark of
Steris Corporation. neodisher is a trademark of Chemische Fabrik Dr. Weigert GmbH & Co. KG. Enzol is
a trademark of Johnson & Johnson.
©2013 Integra LifeSciences Corporation. All rights reserved.
Page 31

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis