Incorrect measurement results.
-
Recognisable consequences of prolonged storage under improper conditions.
-
Recognisable consequences of extraordinary transportation stress.
In such cases the BENNING PV 3 must be switched off immediately, disconnected from the measuring
points and secured to prevent further utilisation.

9.3 Cleaning

Clean the casing externally with a clean dry cloth (exception: special cleaning wipers). Avoid using solvents
and/ or scouring agents for cleaning the instrument. It is important to make sure that the battery compart-
ment and battery contacts are not contaminated by leaking electrolyte.
If electrolyte contamination or white deposits are present in the region of the batteries or battery casing,
clean them too with a dry cloth.

9.4 Battery replacement

c
Before opening the BENNING PV 3 ensure that all the test leads have been disconnected
from the BENNING PV 3. Danger of electric shock!
The BENNING PV 3 is supplied by means of three rechargeable 3.7 V 18650 lithium-ion batteries (accu-
mulators, protected version).
A battery replacement is required, if the battery symbol I appears in the display 1 .
Proceed as follows to replace the batteries:
-
Switch the BENNING PV 3 off.
-
Disconnect all measuring leads from the device.
-
Unscrew the two screws from the battery compartment cover.
-
Lift off the battery compartment cover from the bottom part of the battery compartment.
-
Remove the discharged batteries from the battery compartment.
-
Then, insert the fully charged batteries into the battery compartment at the provided places (please
observe correct polarity of the batteries).
-
Lock the battery compartment cover into place on the bottom part and tighten the screws.
Make your contribution to environmental protection! Do not dispose of discharged bat-
teries in the household garbage. Instead, take them to a collecting point for discharged
batteries and special waste material. Please inform yourself in your community.

9.5 Calibration

Benning guarantees compliance with the technical and accuracy specifications stated in the operating
manual for the first 12 months after the delivery date. To maintain the specified precision of the measure-
ment results, the instrument must be recalibrated at regular intervals by our factory service. We recom-
mend a recalibration interval of one year. Send the appliance to the following address:
Benning Elektrotechnik & Elektronik GmbH & Co. KG
Service Center
Robert-Bosch-Str. 20
D – 46397 Bocholt
Technical Support/ Helpdesk:
Phone: +49 (0) 2871 93-555
Telefax: +49 (0) 2871 93-417
E-Mail: helpdesk@benning.de

9.6 Spare parts

-
Set of measuring leads with probe tip (L = 1.2 m) (red/black) incl. set of alligator clips (red/black), P.no.
10208356
-
Set of PV measuring leads for MC4 connector (L = 1.2 m) (red/ black), P.no. 10208355
-
Three rechargeable 3.7 V 18650 lithium-ion batteries (accumulators, protected version), P.no.
10208358
-
Charger, P.No. 10208360
-
USB connecting cable (A plug to B plug), P.no. 10000315