Press the
-key 6 again to go to the next storage location.
8.7.3
Deleting the measured value memory
-
To delete the entire measured value memory, press the
8.7.4
Warning in case of the measured value deviating by 5 %
-
In case the displayed measured value memory (open-circuit voltage and short-circuit cur-
rent) deviates by 5 % from the average value of all stored measured values, the "5%" symbol
starts flashing.
8.7.5
Average value of the entire measured value memory regarding the open-circuit voltage
and short-circuit current
-
Press the
-key 6 repeatedly until the storage location 0 N is shown on the display 1 .
The symbol "RECALL" M is displayed 1 .
-
The storage location 0 N calculates the average values for open-circuit voltage and short-
circuit current from all stored measured values.
9.
Maintenance
Before opening the BENNING PV 1, make quite sure that it is voltage free!
Electrical danger!
Work on the opened BENNING PV 1 under voltage may be carried out only by skilled
electricians with special precautions for the prevention of accidents.
Make the BENNING PV 1 voltage free as follows before opening the instrument:
-
Switch the tester off.
-
Remove all connecting cables from the object.
9.1 Error codes
Error code
9.2 Securing the instrument
Under certain circumstances safe operation of the BENNING PV 1 is no longer ensured, for
example in the case of:
-
Visible damage of the casing.
-
Incorrect measurement results.
-
Recognisable consequences of prolonged storage under improper conditions.
-
Recognisable consequences of extraordinary transportation stress.
In such cases the BENNING PV 1 must be switched off immediately, disconnected from the
measuring points and secured to prevent further utilisation.
09/ 2011
etc.
or
BENNING PV 1
Remedy
Internal fuse defective, see chapter 9.5 "Fuse re-
placement"
The electronic components of the BENNING PV 1
have reached the maximum admissible temperature.
Disconnect the BENNING PV 1 from the object to be
measured and let it cool down.
The DC short-circuit current exceeds the maximum
value of 10 A. Measurement has been stopped.
The BENNING PV 1 is not correctly calibrated, see
chapter 9.6 "Calibration".
Please return the BENNING PV 1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
Please return the BENNING PV 1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
Please return the BENNING PV 1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
Please return the BENNING PV 1 to an authorized
service center, see chapter 9.6 "Calibration" for the
address.
-key 7 and the
-key 6 .
18