Depth:
Weight:
Shipping Weight:
Operating Ambient
Temperature:
Exhaust Cooling Air
Temperature:
Oxygen Output
Temperature:
Cabinet:
No AP/APG
Standards and Regulatory
Listing:
CE marked model
Electrical:
Placement:
Tubing:
Time of Operation:
Recommended Storage
and Shipping Temperature:
Environment:
Part No 1163145
SECTION 5—TYPICAL PRODUCT PARAMETERS
30.5 cm ± 1 cm (12 in ± 3/8 in)
20.5 kg ± 1 kg (45 lbs ± 2 lbs)
22.7 kg ± 1 kg (50 lbs ± 2 lbs)
10°C - 35°C (50°F - 95°F) at 20-60% relative
humidity
Less than Ambient +19°C (+ 35°F)
Less than Ambient +3°C(+ 6°F)
Impact Resistant flame-retardant plastic cabinet
that conforms to UL 94-V0.
Not suitable for use in the presence of a
flammable anaesthetic mixture.
IRC5PO2AW/IRC5PAW
IEC/EN61000-3-2
IEC/EN61000-3-3,
IEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2
IRC5PO2AW Models Only
ISO8359
MDD 93/42/EEC, Annex I and IX
IRC5PO2AW
No extension cords.
No closer than 30.5 cm (12 in) from any wall,
furniture, draperies, or furniture to assure
sufficient air flow. Avoid deep pile carpets and
heaters, radiators or hot air registers. Floor
location only.
2 m (7 ft) cannula with a maximum 15 m (50 ft)
of Crush-Proof Tubing (DO NOT pinch).
Up to 24 hours per day.
-29°C to 65°C (-20°F to 150°F) at 15-95%
relative humidity.
Smoke and soot-free. No confined spaces
(Example: No closets).
15
Perfecto
™Series
2