:  :   guideline and manufacturer's   
declaration of conformity

electromagnetic immunity

The BEMER SET CLASSIC / PRO is designed for operation in the electromagnetic environment listed below.  
The customer or user of the BEMER SETS CLASSIC / PRO should ensure that the device is used in such an environment. 
EMISSION TESTS
Electrostatic discharge according to IEC 61000-4-2 
Fast transient electrical interferences according to IEC 61000-4-4
Surge voltage / surges according to IEC 61000-4-5
Voltage dips, interruptions and variations of the power supply according to 
IEC 61000-4-11
Power frequency (50 / 60 HZ) magnetic field according to IEC 61000-4-8
Note: U
 is the alternating power supply voltage prior to application of the test level
T
140
IEC 60601 TEST LEVEL
± 6 kV Contact discharge
± 8 kV Air discharge
± 2 kV for power supply lines 
± 1 kV Voltage outer conductor – outer conductor
± 2 kV Voltage outerconductor – earth
< 5% U
 for 1/2 cycles (> 95 % dip in U
)
T
T
< 40 % U
 for 10 cycles (60 % dip in U
)
T
T
<70 % U
 for 25 cycles (30 % dip in U
)
T
T
< 5 % U
 5 s (> 95 % dip in U
)
T
T
3 A/m
COMPLIANCE LEVEL
± 6 kV Contact discharge
± 8 kV Air discharge
± 2 kV for power supply lines 
± 1 kV Voltage outer conductor – outer conductor
± 2 kV Voltage outerconductor – earth
< 5 % U
 for 1/2 cycles (> 95 % dip in U
)
T
T
< 40 % U
 for 5 cycles (60 % dip in U
)
T
T
<70 % U
 for 25 cycles (30 % dip in U
)
T
T
< 5 % U
 5 s (> 95 % dip in U
)
T
T
3 A/m