3.2 Assembling the Mast to the Base

WARNING!
– Use only Invacare parts in the assembly of
this patient lift. The base legs, the mast,
boom, pump assembly and swivel bar are
manufactured to specifications that assure
correct alignment of all parts for safe functional
operation.
DETAIL A
A Hex Bolt
B Washer
The mast assembly may be removed from the base
for storage or transporting. The mast assembly MUST
be properly secured to the base assembly before use.
14
DETAIL B
C Nut
D Mounting
Bracket
1. Put the base on the floor.
Make sure all four casters make contact with
the floor.
2. Lock both rear casters Refer to Detail A.
3. Remove the hex bolt, washers and nut that are located
in the U-shape cut-out of the base Refer to Detail B.
4. Lift the mast to an upright position.
5. Lower the mast onto the mounting bracket.
6. Attach the mast to the base with the hex bolt, washers
and nut. Tighten securely.

3.3 Preparing the Lift for Use

Check and tighten all hardware BEFORE use.