Herunterladen Inhalt Inhalt Diese Seite drucken

Using A Power Cord To Charge The Battery - Invacare Stand Assist Rps350–1Fr Gebrauchsanweisung

Aufstehlifter
Vorschau ausblenden Andere Handbücher für Stand Assist RPS350–1FR:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 45
Invacare® Stand Assist

4.7.2 Using a Power Cord to Charge the Battery

This procedure only applies to the RPS350–1FR.
A = Power Cord Connects Here
22
CAUTION!
– The emergency stop MUST NOT be activated -
otherwise it will be impossible to charge the
battery. While charging takes place, the patient
lift cannot be used. DO NOT attempt to move
the patient lift without unplugging the power
cord from the wall outlet. DO NOT attempt
to use the patient lift if the battery housing is
damaged. Replace a damaged battery housing
before any further use.
1. Attach the power cord to the control box.
2. Plug the power cord into a power outlet.
The battery will charge in approximately 4 hours.
Charging must be done in a room with good air
ventilation.
3. Disconnect the power cord from the power outlet after
the battery has been fully charged.
1078985-I
Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Stand assist rps350–1e

Inhaltsverzeichnis