Information about DIN EN 62353 (VDE 0751-1:2008-8)
Device leakage current
With devices where you cannot determine if all mains affected components
are tested with the replacement device leakage current measurement or the
replacement device leakage current measurement for another reason can
not be carried out, then the device leakage current or the differential current
measurement can be measured.
Allowable values from long-term flow device
leakage current
Application part
Device leakage current generally
Leakage current for devices according to notes 1
and 3
Leakage current for devices according to note 2
Device leakage current for protection class-II de-
vices and protection class-I devices, which are not
connected to the PE conductor conductive com-
ponents.
Patient leakage current
Allowable values from long-term flow equiva-
lent leakage current
Application part
Patient leakage current
72
Direct current
Alternating current
Value in mA
B
BF
CF
0.5
0.5
0.5
2.5
2.5
2.5
5.0
5.0
5.0
0.1
0.1
0.1
Value in mA
B
BF
CF
0.01
–
–
0.10
–
–
®
Testboy
TV 470