Equivalent patient leakage current

Allowable values from long-term flow equiva-
lent patient leakage current
Application part
Equivalent patient leakage current.
Note 3: Mobile X-ray machines with mineral insulation.
Important notes
Equivalent device leakage current insulation test is only valid when all
|
circuits in the device are activated.
The application part type determines the limit value.
|
If all safety-relevant parts are not reached with the equivalent device leakage
current, then a leakage current measurement should be made instead. Use
either the direct method or the indirect method. With the indirect method,
the DUT has to be in an isolated position.
®
Testboy
TV 470
Information about DIN EN 62353 (VDE 0751-1:2008-8)
Value in mA
B
BF
CF
5.0
0.05
71