Leakage current measurement (I-ABL)

Illustration 7
Connect the DUT to the test socket. The DUT has to be switched on in order
to be tested correctly.
With the test lead connected to socket I the following should now be tested:
Protection class-I device: All non-earthed components, that are ex-
|
posed and conductive.
Protection class-II device: All exposed conductive parts.
|
®
Testboy
TV 470
WARNING
Never carry out these measurements until the PE conductor
resistance and the insulation resistance measurements have
been successfully carried out!
The DUT should be connected to the mains voltage and in op-
eration during testing. Therefore, be careful with devices that
emit heat or contain moving parts!