SECTION 8—TROUBLESHOOTING GUIDE
LCD DISPLAY:
DESCRIPTION:
A stuck button is being detected upon
power up.
ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart. Call your
provider and report the problem if alarm continues.
LCD DISPLAY:
DESCRIPTION:
Remaining battery capacity has fallen to
15%. Battery requires charging.
Battery gauge is empty.
Loss of power is imminent.
ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or
revert to normal operating screens
Invacare SOLO
2
INDICATORS:
Continuous audible beep every half
second
RED indicator On.
GREEN indicator is Off.
SOLUTIONS:
1. With the unit turned Off, press each
control button looking for a stuck button.
2. Turn On the unit while ensuring that you
are not continuing to hold the Power
On/Off button once the indicator lights and
LCD begin to illuminate.
3. ONLY press the On/Off button during
startup.
INDICATORS:
Triple audible beep every 10 sec
YELLOW indicator Flashing.
SOLUTIONS:
1. Connect transportable oxygen
concentrator to either AC or DC power,
or use an alternate charged battery
module.
2. If no other power source is available,
Change to another source of oxygen if
alternate power is not available.