NO CHARGING SYMBOL SHOWN ON STANDBY SCREEN
SYMPTOM
With the unit turned
Off and AC or DC
power plugged into
the unit, no charging
icon is displayed.
UNIT IS BEEPING WITH NO LIGHTS OR DISPLAY
SYMPTOM
All power sources are
missing.
Part No 1163190
SECTION 8—TROUBLESHOOTING GUIDE
PROBABLE CAUSE
Power source is no good, or
there is a loose connection.
Battery is outside the allowed
temperature range for charging.
Battery module is not fully
seated.
Standby screen is not being
displayed.
PROBABLE CAUSE
Battery pack has been removed
and there is no external power
While operating without a
battery, the external power is
lost.
35
SOLUTION
Try another power outlet
and check connections to
unit.
Allow unit to cool down to
less than 95° F
(35° C), or warm up to 50° F
(10° C).
Ensure the battery module is
completely inserted.
Press the On/Off button to
display the Standby screen
when using DC power.
SOLUTION
Install any battery pack
available.
Beeping will last 10-15
minutes if no power source
is provided.