100.0 µF
1.000 mF
10.00 mF

7.10 Frequency ranges

Overload protection: 1000 V
Measuring range
100.00 Hz
1000.0 Hz
10.000 kHz
100.00 kHz
Minimum frequency: 1 Hz
Minimum sensitivity: > 5 V
7.11 Temperature ranges °C/ °F (BENNING MM 6-1)
Overload protection: 1000 V
Measuring range
-40 °C - +400 °C
-40 °F - +752 °F
*
The measuring accuracy of the K-type temperature sensor has to be added
to the specified measuring accuracy.
Wire temperature sensor (type K):
The measuring accuracy applies to stable ambient temperatures < ± 1 °C. After
a change of the ambient temperature of ± 2 °C, the measuring accuracy data
will apply after 2 hours.
7.12 PEAK HOLD for AC V/ AC A
Please add ± 150 digits to the specified measuring accuracy.
Square-wave signals are not specified.
8.

Measuring with the BENNING MM 6-1/ MM 6-2

8.1 Preparation for measuring
Store and use the BENNING MM 6-1/ MM 6-2 only under the correct tempera-
ture conditions stated. Always avoid longer exposure to sunlight.
-
Check the rated voltage and rated current stated on the safety measuring
leads. The safety measuring leads supplied with the unit are suitable for the
rated voltage and current of the BENNING MM 6-1/ MM 6-2.
-
Check the insulation of the safety measuring leads. If the insulation is dam-
aged in any way, do not use the leads.
-
Check the continuity of the safety measuring leads. If the conductor in the
safety measuring lead is interrupted, do not use the leads.
-
Before selecting another function with the rotating switch  , always discon-
nect the safety measuring leads from the measuring point.
-
Sources of strong current in the vicinity of the BENNING MM 6-1/ MM 6-2
may cause unstable or incorrect readings.
8.2 Voltage and current measurement
Always observe the maximum voltage to earth potential!

Electrical hazard!

The maximum voltage which may be applied to the sockets
-
COM socket 
-
socket (+) for V, Ω,
 (BENNING MM 6-2)
-
socket for 10 A range L (BENNING MM 6-2)
of the BENNING MM 6-1/ MM 6-2 to earth is 1000 V CAT III.
8.2.1
Voltage measurement
-
With the rotating switch  , select the desired function ( ,
BENNING MM 6-1/ MM 6-2.
-
Connect the black safety measuring lead to the COM socket  on the
BENNING MM 6-1/ MM 6-2.
02/ 2016
110.0 µF
0.1 µF
1.100 mF
0.001 mF
11.00 mF
0.01 mF
AC/ DC
OL indication
Resolution
100.00 Hz
1000.0 Hz
10.000 kHz
0.001 kHz
100.00 kHz
for V
eff
AC
> 20 V
for V
eff
to 100 kHz)
> 0,6 A
for A
eff
AC/ DC
OL indication
- 44 °C - +440 °C
- 44 °F - +824 °F
, Hz, μA,
BENNING MM 6-1/ 6-2
± (1.9 % of reading + 5 digits)
± (1.9 % of reading + 5 digits)
± (1.9 % of reading + 5 digits)
0.01 Hz
± (0.1 % of reading + 5 digits)
0.1 Hz
± (0.1 % of reading + 5 digits)
± (0.1 % of reading + 5 digits)
0.01 kHz
± (0.1 % of reading + 5 digits)
(1 Hz - 10 kHz)
(10 kHz - 50 kHz), not specified for (50 kHz
AC
AC
Resolution
0.1 °C
0.1 °F
Measuring range: - 60 °C to 200 °C
Measurement accuracy: ± 2 °C
(BENNING MM 6-1) or for V, Ω,
Measurement accuracy
Measurement accuracy*
± (1 % of reading + 20 digits)
± (1 % of reading + 36 digits)
,
, Hz
) on the
26