9. Alarm display/output

LCD display
Fifth digit is F
Value for axis is flashing.
Note
If the alarm is again displayed/output after solving the problem
Has the measuring probe or its spindle received some strong shock?
Try replacing with the measuring probe used for the other channel.
74
Item
Terminals
Alarm terminal is
All "H"
Measuring probe
"H"
is not connected or
c o n n e c t i o n h a s
been cut.
Acceleration over
Signal disorder or
connection cut
Overflow
True logic: fifth
digits are all "H".
False logic: fifth
digits are all "L".
All "H"
Alarm terminal is
Initialization is
"H"
required.
Output
BCD
RS-232C
∗E
C R L F
output.
(∗ is A or B.)
∗MFX.XXX
C R L F
output.
(∗ is A or B, M is a space
or a minus sign.)
∗E
C R L F
output.
(∗ is A or B.)
Cause
Solution
The measuring probe was
Reset
exchanged with the power on.
Measuring probe is not
F i x p r o b l e m a n d
connected or connection
reset.
has been cut
The measuring probe's
Reset
s p i n d l e e x c e e d e d t h e
maximum acceleration.
Abnormality in measuring
F i x p r o b l e m a n d
probe's output signal or
reset.
connection has been cut.
Value exceeded 5 digits.
Limit input to 5 digits
and reset.
Move the measuring
Initialization is required to
unit at least 1 mm
calibrate the new pairing of
and reset.
both device and measuring
probe.