Herunterladen Inhalt Inhalt Diese Seite drucken

Sony LT10-105 Bedienungsanleitung Seite 72

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Note
• The display unit's least significant digit (rightmost digit) is
the first digit.
The letters in parentheses have the following meanings
A: 1, B: 2, C: 4, D: 8.
• Pins @¢ and @∞ are connected to the internal circuit for
possible future expansion. Do not make any connections
to them.
BCD output
(See P65 "7-2. I/O timing".)
When DRQ has been received from at #º, and when the
READY output at #¡ goes "L" (ON), the BCD data is output.
• Output logic
True logic or false logic can be selected.
(See P53 "5-1-2. BCD model")
True logic : "L" (ON) is "0".
"H" (OFF) is "1".
• Output format
Whether to hold the BCD output data or to assume
high-impedance when there is no DRQ signal input can
be selected.
(See P53 "5-1-2. BCD model")
64
1st digit
Measuring mode selection input
Current value, maximum value, mimimum value, peak-to
peak value can be selected.
@¡pin (MOD 0)
Measuring mode
Current value
L
Maximum value
H
Minimum value
L
P-P values
H
According to the
×
key switch settings
×
: Either setting possible
SIGN output
Tells whether the output data is positive or negative.
With true logic "H" (OFF) is "–", and "L" (ON) is "+".
Start input
When the peak-hold function has been chosen the "L" (ON)
input will cause the maximum value and minimum value, to
be come the current value (peak-to-peak value=0) and
restart their storing.
Reset input
"L" (ON) sets the measured value to zero.
When there is a preset value this is recalled.
Note
Even when the "L" level is held, the Go/No Go output
(terminals) and the display are not held.
@™pin (MOD 1)
@™
@™
@£ @£pin (M-VALID)
@£ @£
@™
@™
L
L
L
H
H
×
H

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis