6. Commissioning, operation

6. Commissioning, operation

6.1 Key function

GB
Input/Output
TC
RTD
°C/°F
Keys 0 - 9
↓ ↑
Arrow keys
CE
Shift

6.2 Wiring diagrams and instructions

6.2.1 Simulating resistances

1. Select the RTD mode.
2. Connect the device to be checked according the figure below.
3. Select the desired area with the RTD key.
4. In the simulation mode under load, the model CEP3000 provides a current of up to 3
mA. If this limit value is exceeded, a warning message is displayed.
16
Switches between the measure and output mode.
Switching between all available kinds of thermocouples and the
direct output of millivolt by pressing the TC (thermocouple) key.
Switching between all available kinds of resistance thermometers
and the direct output in ohms by pressing the RTD key.
Switching between °C and °F
Enter an output value with these keys. Example: For an output of
20 mV, select millivolt as output range, then press key 2 and key 0
as well as the ENTER key.
The up and down arrow key can be used for small modifications
of an initial value. Press the up or down arrow key for setting the
desired value.
Reset the value by pressing the delete key (CE), then press the
enter key.
Selection of the sub-functions (green lettering on the keys).
WIKA operating instructions, model CEP3000