3. Specifications
Approvals and certificates
CE conformity
EMC directive
GB
Certificate
Calibration
Input and output
signals
Voltage signals in mV
Thermocouples
Type J
Type K
Type T
Type E
Type R
Type S
Type B
Type L
Type U
Type N
10
2004/108/EC, EN 61326 emission (group 1, class B) and
immunity (portable test and measuring equipment)
3.1 calibration certificate per DIN EN 10204
optional: DKD/DAkkS calibration certificate
Measuring
range
-10.00 ... +75.00 mV
-200.0 ... 0.0 °C
0.0 ... 1,200.0 °C
-200.0 ... 0.0 °C
0.0 ... 1,370.0 °C
-200.0 ... 0.0 °C
0.0 ... 400.0 °C
-200.0 ... 0.0 °C
0.0 ... 950.0 °C
-20 ... 0 °C
0 ... 500 °C
500 ... 1,750 °C
-20 ... 0 °C
0 ... 500 °C
500 ... 1,750 °C
600 ... 800 °C
800 ... 1,000 °C
1,000 ... 1,800 °C
-200.0 ... 0.0 °C
0.0 ... 900.0 °C
-200.0 ... 0.0 °C
0.0 ... 400.0 °C
-200.0 ... 0.0 °C
0.0 ... 1,300.0 °C
WIKA operating instructions, model CEP3000
Absolute measurement
uncertainty (all errors
incl.)
0.015 % of reading ±20 µV
With cold junction compensation
0.6 °C
0.4 °C
0.8 °C
0.5 °C
0.8 °C
0.5 °C
0.5 °C
0.4 °C
2.4 °C
1.7 °C
1.3 °C
2.4 °C
1.7 °C
1.4 °C
2.1 °C
1.7 °C
1.3 °C
0.45 °C
0.4 °C
0.7 °C
0.45 °C
1.1 °C
0.6 °C