2.3.5

Product-Specific Data

Manual input range
upstream system volume 0
–100000 ml
*1)
When 0 is input, the upstream
system volume is measured.
Measuring Ranges
– Test pressure: 100–8000 mbar
– Pressure drop: 1–2000 mbar
– Max. measur-
able volume: 100000 ml
– Diffusion:
0.1–999.9 ml/min
– Water
intrusion:
0.1–999 ml/10 min
– Water flow:
0.01–99.9 ml/min
– Bubble point: 100–8000 mbar
Measuring Accuracy
± 0.1% ± 9.5 mbar
– Pressure
– Pressure drop ± 1 mbar
– Net volume
measurement ± 4%
– Diffusion
± 5%
– Water
± 5%
intrusion
– Bubble point ± 50 mbar
Operating Conditions
– Room
temperature
+15°C to +35°C
– Rel. humidity 10-80%
Display
– Size 8.4"
– Resolution 640 + 480 pixels
Data Output
– Service TU RS232|RS485
– Service MU RS232
8
2.3.6

Test Methods

*1)
Test method
Press. drop test:
Diffusion test:
Bubble point test:
Multipoint diffusion
test:
Diffusion and
bubble point test
(complete test)
Intrusion test,
water flow test
2.3.7

General Data

Ambient conditions
Ambient
temperature:
Noise level:
Humidity:
The stated values apply from sea
level up to 2000 m.
!
The unit should be set up under
ambient conditions that produce
no negative effects during opera-
tion and processing, e.g. from
contaminated airflows, metallic
dust or fluids from leaks, conden-
sation, aerosols or widely changing
temperature conditions.
Without ext. ref. tank
All filter systems up to net volume of 100 l
Cartridge systems up to
net volume of 9000 ml
Small systems and cartridge
systems up to net volume
of 9000 ml
Cartridge systems up to
net volume of 9000 ml
Cartridge systems up to
net volume of 9000 ml
Cartridge systems up to
net volume of 9000 ml
+5 up to
+35°C
< 70 dB
95% at 25°C
With ext. ref. tank
Cartridge systems up
net volume of 100000 ml
Small systems and cartridge
systems up to net volume
of 50000 ml
Cartridge systems up to
net volume of 50000 ml