Functionality

Fig. 6-1 | Overview
Sartocheck
®
3 plus
1. printer
2. display
3. keypad
4. LED status lights
1
6.1

General Functional Description

The Sartocheck® 3 plus is a
qualifiable filter integrity tester and
consists of a management unit and
a test unit.
The test unit is a mechanically,
pneumatically and electrically
closed functional unit.
A test unit cannot be operated
directly by the user, but is 100%
controlled by the data management
unit via an internal RS 485 inter-
face. With the help of its electrical
and pneumatic peripheral devices
the data management unit performs
the actual filter tests along with a
series of other additional functions.
Filter Integrity Test
The main purpose of Sartocheck
plus is to test the integrity of mem-
brane filters according to various
methods.
6.2

Test Programs

In the test program, the procedural
steps and the parameters are set
down for the filter integrity test.
A programmed test program can
be loaded into the test unit and
the test will be run independently.
2
– Loading a test program
– Run a test program
The test unit runs through the
3
previously loaded test program, in
other words, it carries out the filter
integrity test and saves the result
in non-volatile memory. The data
management unit monitors the
progress of the filter test, i.e. it
4
reads in the current data in order
to present them to the user.
6.3

Management of the Test Results

In the management unit of the
Sartocheck
®
raw date (the test results) are
processed further, i.e. the units are
recalculated and the numbers are
converted into text.
6.4

Functionality Test

With the function test, the user
is able to interactively test the
functionality of pneumatic
components. The proportional
valves are not explicitly, but
rather implicitly tested in this
context.
6.5

Cleaning

Sartocheck
®
procedure for cleaning and flushing
the internal lines and the volume
reference tank.
®