7.13
Program Database (F4 – Main Menu)
All test programs created are saved in the program database.
Sartocheck
®
3 plus features the option to archive 250 user-programmable test
programs in its internal database. The program data in the program item "Program
database (F4)" can be managed.
The following functions can be selected:
F1:
Load test programs from internal memory
All of the programs available in the Sartocheck
F2:
Upload test program from SD card
You have the option of installing new programs in the Sartocheck
(e.g. transferring from one Sartocheck
F3:
Save backup to SD card
A copy of the program settings is saved to the SD card.
F4:
Load backup from SD card
A copy of the test program settings is loaded from an SD card.
With F4 the test programs are stored in backup format.
This applies to the entire program database only. Individual programs cannot
be stored in backup format.

7.13.1 Test Program From Internal Memory (F1 - program database)

When loading test programs from internal memory, all test programs are displayed
in the program selection list. You can highlight the programs by entering the list
number (F1) in the fields provided for this purpose or via the function keys F2–F10
mark. There is an entry line "From ## to ##"above the program list. In this line,
you can highlight test program blocks (via F1).
Confirm your selection by pressing enter.
The highlighting can be removed with the button "Undo marking".
The highlighted test program is displayed above the function keys F1-F4.
When test program blocks are highlighted, you can scroll through the selection
list with LIST UP|DOWN
You can select the following functions:
F1:
View
Display the parameters predefined in the test program along with
previously defined record data.
F2:
Print
Print out the test program on the internal printer
F3:
Save to SD card
The test program is saved to the SD card
F4:
Delete
Delete the test program
®
3 plus are displayed.
®
®
3 plus to another).
3 plus