Fig. 7-40 | Visualize
Fig. 7-41 | Special functions
Abb. 7-42 | Function test MU
Abb. 7-43 | Special functions:
Function test TU
7.14
Visualization (F3 – Main Menu)
Generally, you can continue to work with the Sartocheck
program can be written) while a filter test is running in the background. If you then
want to get information about the current status of the running filter test, you can
display the current filter integrity test with the "Visualization" function.
7.15
Special Functions (F6 – Main Menu)
In the menu "Special functions" you will find Sartocheck
not directly linked to filter integrity testing.
Here, the function tests for the management unit and the test unit are carried out.
The cleaning procedure and drying procedure for the internal pneumatic block and
internal volume reference tank can be selected here. Furthermore, pressure gauge
functions can be activated in the special functions menu. The user can change
his/her own password as long as he|she has access rights to this menu. For more
details, refer to the menu "Service user management".
In the menu Special functions, the corresponding functions can be activated
with the F1-F6 keys.
7.15.1 Function Test of Management Unit (F1 – special functions)
With the function test MU (management unit), you can test the peripherals of the
Sartocheck® 3 plus.
F1:
Printer function test
Print out all print characters in black and red on the internal printer
F2:
Function test SD card
After inserting an SD card, you can select three different test procedures:
Read only|Check file format|Write and read
F4:
Format SD card
Format SD cards
7.15.2 Function Test of Test Unit (F2 – special functions)
The TU Function test will check the pneumatic block. Here you are interactively
guided through a test menu in which all valves and outlet ports are individually
checked. Please follow instructions in the Sartocheck
diagram you will be able to identify immediately which valves are working properly.
The indicator lights next to the valve will light up green. In the event of an error
they will light up red.
®