Maintenance|Service

8.1

Maintenance

Maintenance Intervals
– Every 6-12 months
To ensure that your Sartocheck
will give you long years of high-
quality, consistent service, you can
have it regularly calibrated. This
is a service we offer as part of
a maintenance agreement. If your
device ever shows irregularities in
the measuring procedure or if you
require information, please contact
one of the representatives of
Sartorius Stedim Biotech or contact
the following location directly:
Sartorius Stedim Biotech GmbH
Service|FT
August-Spindler-Strasse 11
37079 Goettingen, Germany
Phone +49.551.308.0
Fax +49.551.308.3289
www.sartorius-stedim.com
If you require technical customer
service or replacement parts, please
have your device identification
ready when contacting one of the
representatives of Sartorius Stedim
Biotech or Sartorius Stedim Biotech
GmbH directly.
46
8.1.1

Replacing the Microfuses

Warning:
Before exchanging the fine-wire
fuses, turn off the unit and unplug
®
it from the wall outlet (mains
supply).
Fuses:
Two 250 V/16 A time-delay fuses
compliant with DIN 41662 IEC
60127-2/III or IEC 127-3.
Remove the fuse socket (see page
13) and pull out the two fuses.
Install the new fuses back in the
connection block. Now plug the
unit back into the wall outlet.
8.1.2
Sartorius Stedim Biotech
Service Menu
The Sartorius Stedim Biotech
Service menu contains all internal
function thresholds and abort
criteria, the improper changing of
which can result in substantial
measurement errors. Therefore, only
a Sartorius Stedim Biotech service
technician or a specially trained
member of your staff should
change the parameters in the
Service Menu.
A Service Menu instructions are
only handed out to specifically
named and authorized Service
technicians who must confirm
receipt.
8.2

Cleaning

For cleaning, use a wet cloth and
wipe afterwards with a dry cloth.
Make sure that no humidity or
moisture enters the device. Do not
use solvents!
!
Follow the shutdown procedure
(see Chapter 4.4) before carrying
out maintenance, cleaning and
repair work!