Herunterladen Inhalt Inhalt Diese Seite drucken

Adjustment - KERN CFS 300-3 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für CFS 300-3:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

14 Adjustment

• Provide required adjustment weight, cf. chapter 1.
The adjustment weight to be used depends on the capacity of the weight /
counting system. Implement adjustment as close to the maximum load as
possible. Info about test weights can be found on the Internet at:
http://www.kern-sohn.com
• Observe stable environmental conditions. A warming up time (see chapter 1)
is required for stabilization.
• In order to avoid errors at the piece quantity determination, both balances
must be adjusted with the same acceleration due to gravity.
In case of non-compliance counting errors will result!
14.1 Models CFS 300-3, CFS 3K-5
 Switch-on balance and during the selftest press
 Use the numeric keys to enter password:
Either input four times zero „0000" as a standard
password, or input the user-defined password (input cf.
parameter „Pin" chapter 13).
 Confirm input by
 Select bulk material or reference balance via
The appeared [] displays the current balance.
When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried out on
both balances
 If necessary, at balance zero display using
the weighing unit [g / kg or lb], which shall be used for
adjustment. The appeared [] displays the current
weighing unit.
Acknowledge with
CFS/CCS-BA-e-1624
Operation
.
.
Display
„Pin"
.
„Pin" „----„
.
„tECH" „LoCAL"
„tECH" „rEmotE"
select
„tECH" „Unit"
61

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cfs 6k0.1Cfs 3k-5

Inhaltsverzeichnis