 Ensure that there are no objects on the weighing pan.
 Wait for stability display (indicator [] atop
then press
 The currently adjusted adjustment weight (e.g. 6 kg) will
be shown.
If applicable, change the indicated weight value with the
numeric keys.
 Acknowledge with
 Carefully place the indicated weight value for adjustment
into the center of the weighing plate on display „LoAd".
 Wait for stability display, then press
 After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will
return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(
CFS/CCS-BA-e-1624
.
.
), repeat adjustment process.
appears),
.