15.1 Models CFS 300-3, CFS 3K-5
 Switch-on balance and during the selftest press
 Use the numeric keys to enter password „9999":
Confirm input by
 Select bulk material or reference balance via
The appeared [] displays the current balance.
When using as counting system, the bulk material
balance as well as the reference balance must be
linearized. Both scales must be linearized.
 If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be used for
linearization. The appeared [] displays the current
weighing unit.
Acknowledge with
 Ensure that there are no objects on the weighing pan.
Wait for stability display (indicator [] atop
expires), then press
 Carefully place first adjustment weight into the center of
the weighing plate on display „LoAd 1".
Wait for stability display, then press
 Carefully place second adjustment weight into the center
of the weighing plate on display „LoAd 2".
Wait for stability display, then press
 When „LoAd 3" is displayed, place the third adjustment
weight carefully in the centre of the weighing plate.
Wait for stability display, then press
CFS/CCS-BA-e-1624
Operation
.
.
.
.
.
.
.
.
Display
„Pin"
„Pin" „----„
„tECH" „LoCAL"
„tECH" „rEmotE"
„tECH" „Unit"
67

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cfs 6k0.1Cfs 3k-5

Inhaltsverzeichnis