Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance And Storage - 3M Ranger 245 Betriebsanleitung

Blut-/flussigkeitswarmesystem
Vorschau ausblenden Andere Handbücher für Ranger 245:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 15

Maintenance and Storage

CAUTION
• Do not immerse the warming unit in cleaning or sterilizing solutions. The unit is not liquid-
proof.
• Do not clean the warming unit with solvents. Damage to the case, label, and internal compo-
nents may result.
Cleaning
1. Disconnect the Ranger warming unit from the power source.
2. Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute
bleach, or cold sterilants. Do not use abrasive materials.
3. Wipe with a dry, soft cloth.
TO CLEAN THE INSIDE OF THE UNIT
The Ranger cleaning tool is intended to clean both heating plates of the warming unit.
CAUTION
• Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit.
• The cleaning tool provides only superficial cleaning–it does not disinfect or sterilize the
interior of the unit.
1. Unplug the warming unit.
2. Unfold the cleaning tool. Wet the foam pads with a nonabrasive solution, such as Alconox*
10
brand detergent.
3. Insert the tool from the back of the unit and pull the tool all the way out from the front.
4. Rinse the tool with water and repeat 3 times.
5. Wipe off the unit to remove excess fluid.
Note: If the cleaning tool cannot be inserted through the slot of the warming unit because of
excess dried-on fluids, send the unit to a biomedical technician.
Do not use abrasive materials or solutions to clean the heater plates.
Storage
Store all components in a cool, dry place when not in use.
Servicing
There are no user-serviceable parts in the Ranger warming unit. All service must be per-
formed by Arizant Healthcare Inc. or an authorized service technician.
Call Arizant Healthcare Customer Service at 800-733-7775 or 952-947-1200
for service information.
*Alconox is a registered trademark of Alconox, Inc.
Ranger Blood/Fluid Warming System
:

Quicklinks ausblenden: