3.2 Troubleshooting

Ion-selective electrodes, Manual
The following problems may occur:
Measured signal is unstable
Sluggish response time ("running" measured value)
Electrode slope too low
Electrode zero point (E(0)) not correct
Reduced linearity range
These problems can occur on account of the reference or the ion-
selective electrode.
Problems with the reference electrode
Is the quantity of the bridge electrolyte enough?
Are there air bubbles in the bridge electrolyte?
Open and clean the ground-joint diaphragm. Is the flow through the
diaphragm OK after cleaning?
Is the electrode cable correctly plugged in?
Is the electrode cable defective?
Problems with the ion-selective crystal membrane electrode
Polish the sensor surface with the polishing set.
Exception: The sensor surface of the F
tooth paste or liquid soap.
Avoid electrode poisons, see also page 40. Electrode poisons must
be precipitated or masked.
Is the electrode cable correctly plugged in?
Is the electrode cable defective?
The sensor surface may be destroyed. Use new electrode.
Problems with the ion-selective polymer membrane electrode
Don't use organic solvents!
Any other components which may destroy or contaminate the sen-
sor surface must be eliminated.
Avoid electrode poisons, see also page 40. Electrode poisons must
be precipitated or masked.
Is the electrode cable correctly plugged in?
Is the electrode cable defective?
The sensor surface may be destroyed. Use new electrode tip or
new electrode.
3 Maintenance
-
electrode is polished with
39