1 Safety Instructions

• Comply with national regulations governing mounting, electrical installation,
commissioning, maintenance and repair procedures.
• Requirements with regard to specialized technical staff for the mounting,
electrical installation, commissioning, maintenance and repair of the
measuring devices:
– trained in instrument safety
– familiar with the individual operation conditions of the devices
– for Ex-certified measuring devices: also trained in explosion protection
• The spare parts set and Installation Instructions are used to replace a faulty unit
with a functioning unit of the same type.
Only use original parts from Endress+Hauser.
• Using the material number and identification information, check whether the
spare part matches the identification label on the measuring device.
• If you have any questions, contact your E+H service organization.
• Modifications to the measuring device are not permitted.
• Follow the Operating Instructions for the device.
• Before removing the device, stop the process and purge the pipe of dangerous
gases and liquids.
• Hot surfaces! Risk of injury! Before commencing work: Allow the system and
measuring device to cool down to a touchable temperature.
• In the case of measuring devices in safety-related applications in accordance
with IEC 61508 or IEC 61511: Commissioning in according with Operating
Instructions after repair. Document the repair procedure.
Endress+Hauser
• In the case of Ex-certified measuring devices: Only open in a de-energized
state (once a delay of 10 minutes has elapsed after switching off the power
supply) or in environments which do not have a potentially explosive
atmosphere.
• Service connector:
– do not connect in potentially explosive atmospheres.
– only connect to Endress+Hauser service devices.
• Only open housing for a brief period. Avoid the penetration of foreign bodies,
moisture or contaminants.
• Replace defective seal/gaskets by original parts from Endress+Hauser only.
• If threads are damaged or defective, the measuring device must be repaired.
• If spacing is reduced or the dielectric strength of the measuring device cannot
be guaranteed during repair work, perform a test on completion of the work
(e.g. high-voltage test in accordance with the manufacturer's instructions).
• Risk of damaging electronic components! Ensure you have a working
environment protected from electrostatic discharge.
• The measuring device is energized. Danger: Risk of electric shock! Only open
the measuring device in a de-energized state.
• Observe the instructions for transporting and returning the device outlined in
the Operating Instructions.
Levelflex FMP5x
2