Installation Instructions
Levelflex FMP5x
Probe with process connection
EA00045F/00/A2/03.11
71146527
Determining the correct spare part set
This method is used merely to determine whether the spare part set is suitable for the measuring device. Other conclusions drawn from the table are not applicable.
1. Read the order code from the nameplate on the measuring device and enter it in the fields below. The relevant features are marked with a star.
2. Select the order code of the spare part set in the table below and compare it, with the ordered spare part.
Levelflex
FMP50
FMP51
FMP52
FMP54
FMP55
FMP56
FMP57
Order code, set original spare part
XPF0002 -
Probe FMP50 with process connection
XPF0004 -
Probe FMP51 with process connection
XPF0006 -
Probe FMP52 with process connection
XPF0010 -
Probe FMP54 with process connection
XPF0012 -
Probe FMP55 with process connection
XPF0014 -
Probe FMP56 with process connection
XPF0016 -
Probe FMP57 with process connection
The features below the grey boxes (> 500) are optional features with single or multiple value.
!
Note!
Use only the identical original Endress+Hauser spare part sets.
Additionally check the order code of the probe rod/rope with the aid of the serial number in the device viewer (www. endress.com/deviceviewer).
We recommend that the Installation Instructions be kept with the packaging label at all times.
Determining approval and authorized personnel category:
Any person who carries out repairs is responsible for safety during the work, the quality of work completed and safety of the device after repair.
Without approval
Approval
Competence level
2, 3
1 = Trained specialist at customer site, 2 = Service technician authorized by Endress+Hauser, 3 = Endress+Hauser (return device to manufacturer)
* = It is recommended that devices with a design approval be repaired by skilled professionals who have been trained by E+H.
10
20
30
40
50
-
*
-
*
-
*
-
*
-
*
-
*
-
*
10
*
*
*
*
*
*
*
With approval
(e.g. IECEx, ATEX, FM, CSA, TIIS, NEPSI)
2, 3
60
90
100
...
540
+
*
*
*
+
*
*
*
+
*
*
+
*
*
*
E*
+
*
*
+
*
*
*
+
*
*
*
60
90
100
540 580 590 600 610
+
*
*
*
*
+
*
*
*
*
+
*
*
*
+
*
*
*
*
*
+
*
*
*
+
*
*
*
*
+
*
*
*
*
...
580
590 600
610
J*,K*
M*
J*,K* L* M* N*,O*
J*,K* L* M* N*,O*
J*,K* L* M* N*,O*
J*,K* L* M* N*,O*
J*,K*
M*
J*,K*
M*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
71146527
...

Werbung

loading

Inhaltszusammenfassung für Endress+Hauser Levelflex FMP5 Serie

  • Seite 1 2, 3 1 = Trained specialist at customer site, 2 = Service technician authorized by Endress+Hauser, 3 = Endress+Hauser (return device to manufacturer) * = It is recommended that devices with a design approval be repaired by skilled professionals who have been trained by E+H.
  • Seite 2: Safety Instructions

    • Replace defective seal/gaskets by original parts from Endress+Hauser only. Only use original parts from Endress+Hauser. • If threads are damaged or defective, the measuring device must be repaired.
  • Seite 3 1.5/2.5 Nm Grafik Elektronik Sonde_FMP5x_1 Sonde_FMP5x_2 Sonde_FMP5x_3 3 Re-assembly 8 mm 1.5/2.5 Nm 0.5 Nm Sonde_FMP5x_4 Sonde_FMP5x_5 Sonde_FMP5x_6 Sonde_FMP5x_7 Re-assembly of all other parts is carried out in reverse order, see chapter 2, e -> a. 0 Nm Sonde_FMP5x_8 Endress+Hauser...
  • Seite 4 If the probe length is identical with the original length of the probe, a probe calibration is not required. If the probe length or probe type is different, perform a new level calibration according to the description in the Operating Manual. Additionally it is necessary to order a new typeplate and fix it on the instrument! Endress+Hauser...
  • Seite 5 2, 3 1 = Ausgebildete Fachkraft des Kunden, 2 = von Endress+Hauser autorisierter Servicetechniker, 3 = Endress+Hauser (Gerät an Hersteller zurücksenden) * = Es wird empfohlen, Reparaturen, welche an Geräten mit Bauartzulassung vorgenommen werden, von ausgebildeten Fachleuten durchführen zu lassen, die von E+H geschult wurden...
  • Seite 6: Sicherheitshinweise

    • Ersatzteilset und Einbauanleitung dienen dazu, eine defekte Einheit gegen oder Verunreinigung vermeiden. eine funktionierende Einheit des gleichen Typs zu ersetzen. • Defekte Dichtungen nur durch Originaldichtungen von Endress+Hauser Nur Originalteile von Endress+Hauser verwenden. ersetzen. • Anhand von Materialnummer und Kennzeichnung prüfen, ob das vorliegende •...
  • Seite 7 Grafik Display3 0.5 Nm 8 mm 1.5/2.5 Nm Grafik Elektronik Sonde_FMP5x_1 Sonde_FMP5x_2 Sonde_FMP5x_3 3 Zusammenbau 8 mm 1.5/2.5 Nm 0.5 Nm Sonde_FMP5x_4 Sonde_FMP5x_5 Sonde_FMP5x_6 Sonde_FMP5x_7 Der weitere Zusammenbau erfolgt in umgekehrter Reihenfolge, siehe Kapitel2, e -> a. 0 Nm Sonde_FMP5x_8 Endress+Hauser...
  • Seite 8 Gerät wieder in den Behälter einbauen. Entspricht die Sondenlänge der Ursprungslänge vor dem Austausch des Sondengewichts, ist ein Sondenabgleich nicht notwendig. Bei abweichender Sondenlänge muss eine neue Füllstandskalibration gemäß der Betriebsanleitung durchgeführt werden. Zusätzlich muss ein neues Typenschild bestellt und am Gerät angebracht werden. Endress+Hauser...