Herunterladen Diese Seite drucken

Endress+Hauser Levelflex FMP5 Serie Einbauanleitung Seite 3

Vorschau ausblenden Andere Handbücher für Levelflex FMP5 Serie:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Probe with process connection
Tool List
3 mm, 5 Nm
8 mm,
4 mm, 15 Nm
#
Warning!
Before dismounting the main electronic module de-energize the instrument (e.g. disconnect the terminals).
The measuring instrument could be damaged, if the energized main electronic module will be removed.
2 Exchange of probe with process connection
a
3 mm
1.5/2.5 Nm
e
T10
0.5 Nm
Grafik Elektronik
3 Re-assembly
a
Sonde_FMP5x_4
e
T10
0 Nm
Sonde_FMP5x_8
Endress+Hauser
18 mm
14 mm, 10/15 Nm
(remote version)
(divisivle rod probes)
b
Grafik Deckel
f
b
Re-assembly of all other parts is carried out in reverse order, see chapter 2, e -> a.
T10
c
Grafik Display1
g
T10
Sonde_FMP5x_1
c
8 mm
1.5/2.5 Nm
T10
0.5 Nm
Sonde_FMP5x_5
T10
(tool is attached)
d
Grafik Display2
h
1.5/2.5 Nm
T10
Sonde_FMP5x_2
d
Sonde_FMP5x_6
Levelflex FMP5x
Grafik Display3
8 mm
Sonde_FMP5x_3
Sonde_FMP5x_7
3

Werbung

loading