3. Technical data

Unit type:
Digital automatic unit with electric pump for
measuring blood pressure on upper arm
Dimensions:
L = 115 mm x W = 150 mm x H = 54 mm
Weight:
315 g not including batteries
Display:
LCD display (liquid crystal display) for measured
values and control displays
Memory:
60 measured values (stored automatically) and
average value (A)
Measurement procedure:
Oscillometric determination of systole, diastole
and pulse
Reference procedure of
Auscultatory measurement
clinical testing:
Inflating pressure:
approx. 10-20 mmHg above systolic blood
pressure
EN-48
Technical information
Technical information
Pressure display range:
0-300 mmHg
Measurement range:
Systolic: 50-250 mmHg
Diastolic: 40-150 mmHg
Pulse: 40-160 pulses/min
Error limits:
Blood pressure measurement: corresponds to EN
1060 Part 3
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5%
Serial number:
The unit bears a serial number SN which provides
clear identification.
Power supply:
Battery type: 4 x 1.5 V round cells alkali
manganese (LR 6) lithium (FR 6)
Durability: over 800 measurements in 2 years
Optional: Stabilised type A1 mains adapter, output
6 VDC, minimum 600 mA
Cuff:
Metal ring cuff for arm circumferences of 23-43
cm, type 2040
Operating conditions:
Ambient temperature 10 to 40 °C, rel. air humidity
up to 85% not condensing, air pressure 700 to
1040 hPa
Storage and transport
Ambient temperature -5 to 50 °C, rel. air humidity
conditions:
up to 85% not condensing,
Automatic switch-off:
approx. 3 minutes after end of measurement
Protection class IP:
IP 20: Protected against solid foreign particles
with a diameter of more than 12.5 mm, no
protection against water.
Expected service life:
5 years
Classification:
Internal power supply by battery
E