Safety & Warnings
Important safety information, see user manual.
Belangrijke veiligheidsinformatie, zie gebruiksaanwijzing.
Information importante relative à la sécurité, voir mode d'emploi
Wichtige Sicherheitsinformationen, siehe Gebrauchsanleitung.
10
Symbols displayed on the unit
Symbolen op het toestel
Symboles sur l'appareil
Symbole auf dem Gerät
!
Ground level
Aardpotentiaal
Potentiel de la terre
Erdpotential
Aps230 ADVANCED PersonalScope

SAFETY and WARNINGS

The PersonalScope is ideally suited for
measurements of category II installations
with pollution degree 1 and using a maxi-
mum of 600V, in accordance with the
IEC1010-1 norm.
Consequently, all measurements should be
avoided in case of polluted or very humid
air. One should also refrain from measuring
conductors or installations that use voltages
that exceed 600Vrms above ground level.
CAT II indicates conformity for measure-
ments of domestic installations.
The maximum input voltage for the connec-
tions of the unit stands at 100Vp (AC+DC)
Do NOT open the enclosure while perfor-
ming measurements.
Remove all test leads before opening the
enclosure in order to avoid electrical shock.
Use the included measuring probes with an
insulated connector when measuring volt-
ages in excess of 30V (PROBE60S).
Do NOT perform measurements unless
the battery compartment is closed. The
battery cover should only be removed
for the calibration of the probe 's X10-
position or to reset the instrument.