Technical Data

12.0 Technical Data
Low Ohm measurement
Measurement range:
Limit value:
Accuracy of limit value:
Open voltage V q :
Test current:
Test time:
Number of tests:
Over load protection:
Insulation resistance
Measurement range:
Limit value:
Accuracy of limit value:
Nominal test voltage V N :
Open voltage V q :
Test current I N :
Short circuit current I k :
Test time:
Over load protection
Number of tests:
Over load protection:
The EUT will discharge automatically after the measurement
Internal mains voltage resistance
Measurement range:
Limit value:
Accuracy of limit value:
Test current:
Load time:
Mains voltage range:
Nominal mains voltage:
Contact voltage limit:
Test time:
Maximum mains voltage fluctuation during the measurement: 1 %
The protective earth conductor must be free of interfering voltage.
40
0,1...1,9 Ω
1,0 Ω
+0%, -30%
5 V...9 V DC
≥ 200 mA (at resistors ≤ 2 Ω)
approx. 1 s
ca. 4000 (at 1 Ω, 1 test - 25 s break)
500 V AC/DC continous
0,1...20 MΩ
1,0 MΩ
+30%, -0%
500 V DC
V N + 50%
≥ 1 mA at U N
< 3 mA
1 ... 3 s
1 ... 10 s (at capacitive load)
ca. 2800 (500 kΩ, 1 test - 25 s break)
500 V AC/DC continuous
600 V AC/DC for 10 seconds
0,3...10 Ω
1,5 Ω
+0%, -30%
ca. 1 A
< 80 ms
195...264 V
230 V/240 V
50 ... 60 Hz, cos ϕ > 0,95
25 V (+0%, -20%)
ca. 3 s