P P L I C A T I O N S

2.1 Intended use

Electrical provision of a continuous vacuum between 0 and -95kPa / 0 to -700mmHg for use in the clinic,
hospital and doctor's medical practice.
The device is designed for continuous operation.
The vacuum provided by the device can be used, depending on the accessories, for example for:
-
Surgical Suction
-
Secretion Suction
-
Vacuum-Extraction
-
Endoscopy
-
Dental and Maxillary Surgery
Please note:
These Operating Instructions only contain general information for the use of the device.
The correct application of the device performance and any accessories
is the responsibility of the medically trained user.

2.2 Indication

The Master/Senator vacuum pump has no medical indication.

2.3 Contraindication

The Master/Senator vacuum pump has no medical contraindication. However, it must not be used in the
following cases:
-
Cardiac surgery
-
Use on the central nervous system
-
Operation with a low, precise vacuum, such as for Thorax drainage
-
Domestic use
-
Outdoor use and in means of transport
-
When moved with mobile power sources

2.4 Notice to the user

In accordance with Regulation (EU) 2017/745 on medical devices, serious incidents that have occurred in
relation to the device should be reported to the manufacturer and the competent authority of the member
state in which the user is established. The required user information is supplied to the customer together
with the product in the form of Operating Instructions. A printed version of the Operating Instructions is
included in the scope of delivery.