5 . C
L E A N I N G
The device and accessories are not disinfected or sterile when delivered.
For cleaning, disinfection or sterilization, the respective Operating Instructions and required cleaning
regulations at the place of use must be observed.
Before cleaning, the device must be switched OFF and the mains power cable must be
disconnected from the mains supply. Failure to comply may result in electric shock!
If the pump unit, and/or accessories such as the foot operated vacuum regulator are
contaminated by failure to use the hydrophobic bacterial filter, by overflow or by another
cause, the contaminated parts must be disposed of by the user in accordance with local
regulations.
Do not use cleaning agents and disinfectants based on phenol!
If disregarded, there is a risk of destruction of the plastic materials by fine cracks.
Do not leave the disinfectant in the suction jar. The plastic material is thereby exposed to
unnecessary chemical action and the strength of the suction jar may be affected.

5.1 Device housing

The surfaces should be cleaned with a damp cloth and a mild detergent.
After cleaning, dry the surfaces with a clean, dry cloth.
For disinfection, only use known agents that do not affect painted surfaces and plastic parts. For
example: perform
®
sterile from Schülke. The recommendations of the disinfectant manufacturer regarding
application and dosage must be followed.

5.2 Suction jar, lid of suction jar and silicone tubing

Ensure that the suction jar, lid and the silicone tubing are cleaned, disinfected and
autoclaved together. This prevents any mix-up of the parts.
-
Remove the bottle from the fitting rail; dispose of bottle content in accordance with local regulations by
twisting off the lid and turning the container upside down.
-
Autoclave to a maximum of 134°C for a maximum of 15 minutes.
Residues of disinfectants must be removed before autoclaving or neutralized.
Only autoclave dry parts.
Do not stack the parts when autoclaving.
-
Thermal disinfection: do not use alkaline detergents.
-
Chemical disinfection: With products that break down the chains of residues and, as a result, dissolve
these residues.
-
Maximum permissible long-term temperature 140°C in steam and 150°C in dry heat.
-
Provided the functional check described in section 4.6 has been passed, the parts can be used and
reprocessed.
Part description
Suction jar
Lid with handle
Couplings connectors green and transparent
Couplings green and transparent
Pressure pin
Clamp with threaded pin
Pressure spring
4x O-ring 8.1 x 1.6mm
O-ring silicone 12.42 x 1.78mm
Connection tubing and Patient tubing
Seal for lid of suction jar
Retainer for float of overflow safety device
Float for overflow safety device
Ardo medical AG
Material
PSU Polysulfone
Grivory® GVX 5H
stainless steel
Silicone
EPDM Ethylene-Propylene-Diene (rubber)
POM Polyacetal (Polyoxymethylene)
PP Polypropylene
15
Operating Instructions
Master and Senator
Subject to modifications