Herunterladen Inhalt Inhalt Diese Seite drucken

Tonometer Prism Field Check - Keeler D-Kat R Gebrauchsanweisung

Inhaltsverzeichnis
5. Dry the disinfected Tonometer prism
with a clean soft non-shedding cloth.
Safely dispose of the disinfectant fluids used.
Do not disinfect using the following: Alcohol, Acetone, UV radiation,
Sterilisation, Immersion in fluid for more than one hour, Temperatures
greater than 60°C.

4. TONOMETER PRISM FIELD CHECK

Check the Tonometer prism under the Slit lamp and ensure there are no
cracks / chips. The chemical used in the diagnosis process (for example:
Fluorescein) will get into the cracks and will show up if observed under
the slit lamp. Do not use it if there is any sign of cracks / chips.
D-KAT Tonometer by KEELER
6. Store the Tonometer prism in a
suitable container ready for use.
7
EN
Inhaltsverzeichnis
loading

Verwandte Produkte für Keeler D-KAT R

Diese Anleitung auch für:

D-kat tD-kat

Inhaltsverzeichnis