D-KAT Tonometer by KEELER

3.2 DISINFECTING THE TONOMETER PRISMS

Always disinfect the Tonometer prisms before use. Hand hygiene must be
considered to prevent any contamination.
1. Carefully remove the Tonometer prism
from the prism holder.
3. Immerse the Tonometer prism in the
disinfectant fluid. Types of disinfectant
fluid vary.
Please follow disinfectant solution guidelines
for instructions, concentration and time of
immersion. (For example: Pantasept – 3%
aqueous solution for 10 minutes, Hydrogen
Peroxide 3% aqueous solution for 10 minutes,
Sodium Hypochlorite, 10% aqueous solution for
10 minutes etc.).
EN
2. Wash the Tonometer prism under cold
4. Rinse the disinfectant from the prism
10 MINS
6
running water for approximately 1
minute, to ensure the Tonometer prism
is physically clean before exposed to
the disinfection process.
in running water for between 10 and
30 minutes.
1 MIN
10-30
MINS