D-KAT Tonometer by KEELER
•
Never use the instrument if the ambient temperature, atmospheric pressure, and / or
relative humidity are outside the limits specified in this manual.
•
Should the instrument suffer shocks (for example, should it accidentally fall), follow the
check procedure outlined in the 'Servicing and Calibrations' (Section 13); if necessary,
return the instrument to the manufacturer for repair.
•
Use only the listed accessories in conjunction with the instrument; use said accessories
only in accordance with the procedures set forth in the instruction manuals.
•
Always carefully observe the safety rules and other precautions published herein.
•
Do not use in the presence of flammable gases / liquids, or in an oxygen rich environment.
•
This device is intended to be used only by suitably trained and authorised healthcare
professionals.
•
This product should not be immersed in fluid.
•
Electrical equipment can be affected by electromagnetic interference. If this occurs whilst
using this equipment, switch the unit off and reposition.
•
If during measurement the slit lamp is moved forward toward the patient or the patient
moves toward the slit lamp, the sensor arm will be pushed into contact with a stop spring
and an audible alarm will sound.
CAUTION
•
Use only genuine Keeler approved parts and accessories or device safety and performance
may be compromised.
•
Keep out of the reach of children.
•
To prevent condensation from forming, allow instrument to come to room temperature
before use.
•
For indoor use only (protect from moisture).
•
There are no user serviceable parts inside. Contact authorised service representative for
further information.
•
Follow guidance on cleaning / routine maintenance to prevent personal injury / damage to
equipment.
•
Failure to carry out recommended routine maintenance as per the instructions in this IFU
may reduce the operational lifetime of the product.
•
At product end of life dispose of in accordance with local environmental guidelines
(WEEE).
EN
4