bipoJet
Mechanical cleaning/disinfecting
Put the ceramics protector on the product.
Place the product on a tray that is suitable for
cleaning (avoid rinsing blind spots).
Store products with hinges or joints on the tray in
such a way that the joints are kept open.
Mechanical cleaning/disinfecting with manual pre-cleaning
Manual pre-cleaning with brush
Stage
Step
I
Disinfecting
cleaning
II
Intermediate rinse
D–W:
Drinking water
RT:
Room Temperature
Stage I
Fully immerse the product in the disinfecting
solution. Make certain that all accessible surfaces
are moistened.
Clean the product with a suitable cleaning brush
(e.g. toothbrush) until all visible residues have been
removed from the surface.
Brush through all surfaces that are not accessible
to visual inspection, e.g. in products with hidden
crevices, lumens or complex geometries, for at
least 5 min or until no more residues can be
removed. Mobilize non-rigid components, e.g. set
screws, links, etc. during cleaning.
After that, thoroughly rinse (at least 5 times) with
the cleaning solution these difficult to access areas
of the product, using a single-use syringe (20 ml).
To avoid the risk of corrosion, do not use a metal
brush or other abrasives that would damage the
surfaces for cleaning the product.
8
Aesculap Surgical Instruments
T
t
[°C/°F]
[min]
RT
15
(cold)
RT
1
(cold)
Conc.
Water
[%]
quality
2
D–W
BBraun Stabimed; aldehyde phenol
and QAV free; pH = 9
-
D–W
-
Stage II
Rinse the product completely (all accessible
surfaces) under running water.
Chemical