Herunterladen Inhalt Inhalt Diese Seite drucken

Sartorius PMA.Evolution EVO1X Betriebsanleitung Seite 49

Farbmischwaagen für den einsatz in explosionsgefährdeten bereichen
Vorschau ausblenden Andere Handbücher für PMA.Evolution EVO1X:
Inhaltsverzeichnis

Werbung

to EU-Type Examination Certificate No. FM16ATEX0029X
Description of Equipment or Protective System:
13
The YCO14-Y Ex-Link Converter is a dedicated device providing power and data to the
PMA.Evolution EVO1X or LAB1X weighing platforms (FM14ATEX0085X). All signal outputs to the
PMA.Evolution are available in RS232 communications. The YCO14-Y Ex-Link Converter is housed
within non-metallic enclosure. A dedicated potential equalization connection is provided on both the
PMA.Evolution and the YCO14-Y Ex-Link Converter. Power and data connections between the
YCO14-Y and the PMA.Evolution are made via a multi-conductor CAT6 cable.
Ratings - The PMA.Evolution and the YCO14-Y Ex-Link Converter are designed to be used together as a
system. The input voltage to the YCO14-Y Ex-Link Converter is via USB connections a 5 V ± 0.25 V. The
ambient operating temperature is 0°C to 40°C.
YCO14-Y "ic" Input Entity Parameters:
Ui = 17.0V, Ii = 2.0A, Ci = 5.7µF, Li = 0mH
Ui = 16.0V, Ii = 2.5A, Ci = 5.7µF, Li = 0mH
YCO14-Y with "nA" Input Circuit (non-sparking apparatus) or outside hazardous locations:
Um = 250V
Specific Conditions of Use:
14
1. The YCO14-Y Ex-Link Converter shall only be used with the PMA.Evolution.
2. The protective ground connectors on the PMA.Evolution and the YCO14-Y Ex-Link Converter shall be
connected together using a potential equalisation conductor with a cross-sectional area of at least 4 mm
3. The CAT6 cable used between the PMA.Evolution and the YCO14-Y Ex-Link Converter shall be
obtained from Sartorius.
4. The enclosure of the YCO14-Y is non-conducting and may generate an ignition-capable level of
electrostatic charges under certain extreme conditions. The user shall ensure that the equipment is not
installed in a location where it may be subjected to external conditions that might cause a build-up of
electrostatic charges on non-conducting surfaces, additionally; cleaning of the equipment should be done
only with a damp cloth.
5. The equipment shall be installed in such a way that it is protected against UV radiation and the entry of
solid foreign objects or water capable of impairing the safety of the apparatus. Reduce the risk of
mechanical damage to a minimum.
6. If the Ex-Link Converter YCO14-Y is installed in Zone 2, the installation must be done in such a way that
the pollution degree 2 of the YCO14-Y board is guaranteed.
7. In Zone 2 the „USB Power" port may not be used.
8. If the cable contains no Ex ic circuits, the cable must be protected against unintended release by use of
Sartorius „Strain Relief for USB Cord" type YSR01.
Essential Health and Safety Requirements:
15
The relevant EHSRs that have not been addressed by the standards listed in this certificate have been
identified and assessed in the confidential report identified in item 8.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals Ltd. 1 Windsor Dials, Windsor, Berkshire, UK. SL4 1RS
T: +44 (0) 1753 750 000 F: +44 (0) 1753 868 700 E-mail:
F ATEX 020 (Apr/16)
SCHEDULE
atex@fmapprovals.com www.fmapprovals.com
Konformität & Lizenzen
Page 2 of 3
Betriebsanleitung EVO1X | LAB1X
2
.
49

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Pma.evolution lab1xPma.hd lab1xPma.hd evo1x

Inhaltsverzeichnis