Brevera
System
®
Setup Guide for Hologic Affirm
Multicare
Platinum Products
®
PLEASE READ PRIOR TO PERFORMING PROCEDURE:
This setup guide is designed to be used in conjunction with, not replace, the Brevera breast biopsy device
instructions for use. Prior to performing the procedure, the physician must review and be familiar with the full
operating instructions for the Brevera breast biopsy device and biopsy image guidance system. Consult the Brevera
breast biopsy device instructions for use for full operational instructions, as well as any contraindications and safety
information.
Please retain this setup guide. Note that while Hologic frequently updates the contents of this guide, medical
information and technology can change rapidly. Therefore some of the information in this guide may be out of date.
Use of these instructions with any product other than those specifically noted may result in injury to the patient
and/or damage to the equipment.
Program the breast biopsy image guidance system for the Brevera breast biopsy device according to the needle
setup instructions provided in the appropriate breast biopsy guidance system user manual. Specific parameters are
listed below.
Hologic Affirm and Multicare Platinum Products are Compatible with the following Brevera Components
and Biopsy Accessories:
Catalog Number
BREVDRV
BREVADPTR
BREV09
BREVDISP09
EVIVA_NG09L
Setup of Hologic Affirm Biopsy Guidance System with or without the
Lateral Arm Upright Biopsy Accessory and Affirm Prone Biopsy System for
Use with Brevera Biopsy
To Install the Brevera Adapter:
1. Align the outer holes in the adapter with the guide pins on the mount.
2. Align the center hole with the mount screw.
Check the Brevera Adapter components (reference Figure 1) to ensure they are secure.
If the components are not secure call technical support.
3. Turn the knob clockwise on the mount to attach the adapter.
4. Place needle guide in needle guide holder.
Check to make sure the cable, tubing, and other parts of the Brevera biopsy device are clear of the x-ray tube
during rotation.
Figure 1: How to Install the Brevera Adapter
English
Description
Brevera Driver
Brevera Adapter
Standard Stereotactic Biopsy Needle
Standard Stereotactic Biopsy Needle
Needle Guide
2
and
®
Figure Legend
1. Adapter
2. Mount Screw
3. Mount Knob
4. Needle Guide Holder