Instruction manual
5.2.2
P01 Output source
FACTORY DEFAULT: 1011
P01ba defines the source of the primary output value (HART – PV), which also defines the value transmitted on the analog current
output. The device automatically selects the measurement mode according to the selected output source. The device measures the
level's distance. The other quantities are calculated based on the specified tank parameters and material characteristics.
P01
ba
Measurement configuration ➙ Measurement mode
EView2
SV
PV
Output data / measuring
'dc'
'ba'
mode
10
Distance
11
Level
12
Volume
13
Weight
14*
Flow
15
Empty Volume
16
Level%
17
Volume%
40
TEMP
41
TOT1
42
TOT2
Distance measurement (DIST)
/ Level measurement (LEV)
DIST: Currently measured distance
A: Shortest measurable distance (P05)
H: Longest measurable distance, it is also the zerolevel
distance (P04)
Parameters
–
P04
P04, P40...45
P04, P32, P40...45
P04, P40...45, P46
P04, P40...45, P47
P04
P04, P40...45
–
–
–
Volume measurement (VOL)
DIST: Currently measured distance
A: Shortest measurable distance
H: Zero-level distance
B: Volume associated with the highest measurable level
C: Tank's total volume
Type 2298 80 GHz Radar Level Sensor
* Cannot be selected in Volume (12, 17), Weight (13),
and Empty Volume (15) measuring modes.
Volume measurement (VOL)
25